Features
• Ideal for low gas concentration (ppb detection)
• Easily replaceable plug-in tape cassette
• Cassette tape remainder display included
Its best application then is in monitoring of the target gas at the outlet of an absorption scrubber, as well as in monitoring of the work environment for personnel safety.
Since the cassette can be replaced so quickly and easily by a one touch cassette method, detection of multiple gases is practical.
Applications
• Environmental Monitoring
• Semiconductor Industry
• Manufacturing
Specifications
Model
FP-260
Type Desktop type, transportable
Detection principle Photoelectric photometry method using paper tape cassette
Detection tape Cassette type, in VHS style quick change cassette
Accuracy Within (+/- 20% of indication value +1 digit) at constant conditions
Tape operation time 1 month (with no alarm)
- Tape remainder display and caution
- Alarm display at tape end
Sampling flow rate Approx. 1.0 l/min
Gas Inlet Outer diameter: 6mm, teflon tube with fitting
Response Time Within 60 seconds when exposed to gas concentration of 1.6 times alarm level.
Display Digital LCD
Alarm Preset Value Warning: 1/6 of full scale
Alarm: 1/3 of full scale
Alarm Display Lamp (warning-yellow, alarm-red) and buzzer
Alarm Output Latching (standard) or non latching (option)
- Dry contact 1A (warning) latching AC 125V, 0.5A
- Dry contact 1A (alarm) latching AC 125V, 0.5A
External output signal 4-20mA, max load resistance 300Ω (standard or Multiplex transmission (option)
Power supply & power 100VAC, 115 VAC, 220VAC 50/60 Hz (35VA at Measurement, 60VA at feeding tape)
Operating temperature & humidity 5~35°C (41~95°F), 30~90% RH (non condensing)
Dimensions & weight Approx. 275(W) x 220(H) x 370(D)mm
Approx. 12 kg
Self diagnosis Sensor trouble, flow fail, warning of tape end, tape exchange or tape damage, system trouble. Lamp and LCD message "C" contact (No voltage, self-diagnosis) rating : AC 125V, 0.5A
Detectable Gases and Detection Ranges
Gas / TLV value / Detection Range / Resolution
Phosphine (PH3) 300 ppb / 0~900 ppb/ 5
Silane (SiH4) / 5 ppb / 0~15 ppm / 0.25
Diborane (B2H6) / 100 ppb / 0~300 ppb / 5
Arsine (AsH3) / 50 ppb / 0~150 ppb / 1
Disilane (Si2H6) / 5 ppm / 0~10 ppm / 0.1
Hydrogen Sulfide (H2S) / 10 ppm / 0~100 ppb /1
Phosgene (COCl2) / 100 ppb / 0~300 ppb / 2
Chlorine (Cl2) / 0.5 ppm / 0~100 ppb / 1
Hydrogen selenide (H2Se) 50 ppb / 0~300 ppb / 2
Ordering Information
FP-260K Model FP-260K paper tape machine
260AGZS Model FP-260AGZS paper tape machine for C4F6 / C5F8
Replacement Parts
65-FV-017 Tape FV-017 for Hexafluorobutadiene C4F6 & C5F8
65-FVL-001 Tape FVL-001 for Cupra Select, Arsine AsH3, Phosphine PH3, Disilane SiH6, Hydrogen Sulfide H2S,Hydrogen Selenide H2Se
65-FVL-002 Tape FVL-002 for Silane SiH4, Diborane B2H6, Dimethylaluminum Hydride DMAH (C2H7AL), Cupra select
65-FVL-004 Tape FVL-004 for Hydrogen Chloride HCl
65-FVL-006 Tape FVL-006 for Hexafluoroathetyle Acetone Hhfac (CF3COCH2COCF3)
65-FVL-015 Tape FVL-015 for Ruthenium tetraoxide RuO4, Chlorine Cl2
65-FVL-108 Tape FVL-108 for Phosgene COCl2
Use the request button on product pages to build a quote request list.